Keysight EXG and MXG X-Series Signal Generators User’s Guide
215
Basic Digital Operation (Option 653/655/656/657)
Using Waveform Markers
Setting Marker Polarity
Setting a negative marker polarity inverts the marker signal.
1.
In second Arb menu (
), press
Marker Utilities
>
Marker Polarity
.
2.
For each marker, set the marker polarity as desired.
—
The default marker polarity is positive.
—
Each marker polarity is set independently.
“Saving Marker Polarity and Routing Settings” on page
As shown on
Positive Polarity:
On
marker points are high (
≈
3.3V).
Negative Polarity:
On
marker points are low (0V).
RF blanking always occurs on the low part of the signal regardless of the polarity setting.
Controlling Markers in a Waveform Sequence
In a waveform segment, an enabled marker point generates an auxiliary output signal that is routed
to the rear panel EVENT output (described in
“Rear Panel Overview (N5171B, N5172B, N5181B, &
) corresponding to that marker number. For a waveform sequence, you
enable or disable markers on a segment–by–segment basis; this enables you to output markers for
some segments in a sequence, but not for others. Unless you change the sequence marker settings
or cycle the power, the marker setting for the last segment edited in the sequence applies to all
segments in the next sequence that you build. For information on building a waveform sequence,
see
.