234
Keysight EXG and MXG X-Series Signal Generators User’s Guide
Basic Digital Operation (Option 653/655/656/657)
Clipping a Waveform
Configuring Rectangular Clipping
Use this example to configure rectangular clipping. Rectangular clipping clips the I and Q data
independently. For more information about rectangular clipping, refer to
Peak–to–Average Power” on page 230
.
Copy a Waveform File
1.
Display the signal generator’s files: Press
File
>
Catalog Type
>
More
>
Volatile Segments
.
2.
Highlight the waveform
RAMP_TEST_WFM
.
3.
Press
Copy File.
4.
Name the copy (in this example, the name is
MY_TEST_REC
) and press
Enter
.
Apply Rectangular Clipping to the Copied Waveform File
1.
Open the DUAL ARB Waveform Utilities menu: Press
Mode
>
Dual ARB
>
More
>
Waveform
Utilities
.
2.
In the list of files, highlight the copied file (in this example,
MY_TEST_REC
).
3.
Create the CCDF plot: Press
Plot CCDF
.
4.
Observe the shape and position of the
waveform’s curve (the dark line in the
example at right).
5.
Activate rectangular clipping: Press
Return
>
Clipping
>
Clipping Type
until
|I|,|Q|
highlights.
6.
Set 80% clipping for the I data: Press
Clip |I| To
>
80
>
%
.
7.
Set 40% clipping for the Q data: Press
Clip |Q| To
>
40
>
%
.
8.
Apply the rectangular clipping to the waveform: Press
Apply to Waveform
.
9.
Create the CCDF plot (see the example
at right): Press
Plot CCDF
.
10.
Observe the waveform’s curve after
clipping.
Note the reduction in peak–to–average
power relative to the previous plot.
Clipping is non–reversible and cumulative. Save a copy of the waveform
file before you apply clipping.
Example waveform curve before clipping
Example waveform curve after rectangular
clipping