Keysight EXG and MXG X-Series Signal Generators User’s Guide
BERT (Option UN7)
Bit Error Rate Tester–Option UN7
In this example, synchronization occurs after receiving a trigger.
The reference data is generated by stored data bits. If the BERT measurement accepts data bits
immediately after receiving a trigger, set the trigger delay to On and the trigger delay count to a
value corresponding to the data format. For PN9 set the delay to 9. Refer to
Figure 12-9