30
Keysight EXG and MXG X-Series Signal Generators User’s Guide
Preferences & Enabling Options
User Preferences
Reference Oscillator Tune
Tunes the internal VCTXCO oscillator frequency.
The user value offsets the factory tuned value (the value is added to the
factory calibrated DAC value). The tune value of 0 sets the factory
calibrated value.
The range; –8192 to 8192, can be set by using the front panel keypad,
knob or remote command.
For details on each key, use key help
as described on
See also the
SCPI Command Reference
.
Utility
>