80
Keysight EXG and MXG X-Series Signal Generators User’s Guide
Optimize Performance
Using the Power Meter Servo
Power Meter Servo Configuration
The following procedure is a basic configuration for using the signal generator’s Power Meter Servo
mode.
1.
Connect the equipment as shown in
Figure 4-15
Power Meter Servo Configuration
2.
Set the signal generator frequency and amplitude to the desired settings for your device.
3.
On the signal generator, press AMPTD > Leveling Control > Power Meter Servo Configuration.
4.
Press Channel A Setup and configure the power meter. Refer to
5.
Press Return > Return > Settling Error. Set the Settling Error value.
6.
Press Return > Return > Leveling Mode. Select the leveling mode.
Power Meter Once performs the adjustment only at the end of any transition (amplitude or
frequency change). After the adjustment is performed and the power is corrected, no further
adjustments are performed until the next transition.
The configuration described below is one possible setup example.
Consider the limits of your DUT and use caution to protect the DUT from
being exposed to too much power.
To get the best results and reduce stress on the DUT, the amplitude offset
should be set to match the approximate gain of the device. For example, if
the DUT is an amplifier with 20 dB of gain, set the amplitude offset to 20
dB.
Press
AMPTD > More > Amptd Offset
, then enter a value.
For details on each key,
use key help as described
on
.
Power Sensor
Directional Coupler
Содержание N5171B EXG
Страница 1: ...User s Guide Keysight X Series Signal Generators N5171B 72B 73B EXG N5181B 82B 83B MXG...
Страница 4: ...iv...
Страница 356: ...340 Keysight EXG and MXG X Series Signal Generators User s Guide BERT Option UN7 Verifying BERT Operation...
Страница 452: ...436 Keysight EXG and MXG X Series Signal Generators User s Guide Troubleshooting Contacting Keysight Technologies...