174
Keysight EXG and MXG X-Series Signal Generators User’s Guide
Pulse Modulation (Options UNW and 320)
Example
Example
The following example uses the factory preset pulse source and delay.
1.
Preset the signal generator.
2.
Set the frequency to 2 GHz.
3.
Set the amplitude to 0 dBm.
4.
Set the pulse period to 100 microseconds: Press
Pulse
>
Pulse Period
>
100
>
usec
.
5.
Set the pulse width to 24 microseconds: Press
Pulse
>
Pulse Width
>
24
>
usec
6.
Turn on both the pulse modulation and the RF output.
The
PULSE
annunciator displays and the RF output LED lights.
If the modulation does not seem to be working properly, refer to
“No Modulation at the RF Output”
Output:
A 2 GHz, 0 dBm carrier modulated by a 24
μ
s pulse that has a period of 100
μ
s.