![Keysight Technologies N5171B EXG Скачать руководство пользователя страница 337](http://html1.mh-extra.com/html/keysight-technologies/n5171b-exg/n5171b-exg_user-manual_1962194337.webp)
Keysight EXG and MXG X-Series Signal Generators User’s Guide 321
Baseband Operating Mode—Primary, BERT, or N5102A
To set the Baseband Operating Mode to N5102A
To set the Baseband Operating Mode to N5102A
In this mode, all installed and licensed signal generator features as well as all
N5102A features are fully supported, but BERT features are not available.
Figure 11-3
Changing to the N5102A Baseband Operating Mode
1.
Press
Aux Fctn
.
2.
Press
Baseband Operating Mode
.
(Default selection is
Primary
, but image above shows changing from
BERT
.)
3.
Press
N5102A
. (Press the
Return
key to cancel the change.)
4.
After reading the above caution and taking any needed actions,
press
Confirm Change to N5102A Operating Mode
or
Return
to cancel.
This confirmation selects the N5102A ARB FPGA image. All N5102A
features are fully supported, but BERT is not available in this N5102A
Baseband Operating Mode. For backwards compatibility, selecting the
N5102A Baseband Operating Mode sets the I/Q Digital Bus Connectivity to
N5102A.
N5102A annunciator is displayed indicating that the
N5102A Baseband Operating Mode is currently selected.
(For more information on annunciators, see
Changing the Baseband Operating Mode selection will cause all files in the
volatile w
aveform memory to be lost. Backup any files you want to keep before
confirming the change.
See
“Working with Files” on page 56
.
Содержание N5171B EXG