
424
Keysight EXG and MXG X-Series Signal Generators User’s Guide
Troubleshooting
Display
Display
The Display is Too Dark to Read
Brightness may be set to minimum. Use the figure in
to locate the
brightness softkey and adjust the value so that you can see the display.
The Display Turns Black when Using USB Media
Removing the USB media when the instrument begins to use it can cause the screen to go black.
Cycle instrument power.
Signal Generator Lock–Up
— Ensure that the signal generator is not in remote mode (the
R
annunciator shows on the display).
To exit remote mode and unlock the front panel, press
Local Cancel/(Esc).
— Ensure that the signal generator is not in local lockout, which prevents front panel operation. For
information on local lockout, refer to the
Programming Guide
.
— If a progress bar appears on the signal generator display, an operation is in progress.
— Preset the signal generator.
— Cycle power on the signal generator.
Содержание N5171B EXG
Страница 1: ...User s Guide Keysight X Series Signal Generators N5171B 72B 73B EXG N5181B 82B 83B MXG...
Страница 4: ...iv...
Страница 356: ...340 Keysight EXG and MXG X Series Signal Generators User s Guide BERT Option UN7 Verifying BERT Operation...
Страница 452: ...436 Keysight EXG and MXG X Series Signal Generators User s Guide Troubleshooting Contacting Keysight Technologies...