88
Keysight EXG and MXG X-Series Signal Generators User’s Guide
Optimize Performance
Using Flatness Correction
c.
Open the User Flatness table editor and preset the cal array:
Press
Return
>
Configure Cal Array
>
More
>
Preset List
>
Confirm Preset with Defaults
.
d.
In the Step Array menu, enter the desired flatness–corrected start and stop frequencies,
and the number of points:
e.
Populate the user flatness correction array with the step array configured in the previous
step:
Press
Return
>
Load Cal Array From Step Array
>
Confirm Load From Step Data
.
f.
Set the output amplitude to 0 dBm.
g.
Turn on the RF output.
Press
More > Configure Step
Array >
Freq Start > 500 > MHz >
Freq Stop > 1 > GHz >
# of Points > 10 > Enter