Keysight EXG and MXG X-Series Signal Generators User’s Guide
87
Optimize Performance
Using Flatness Correction
4.
Enable the power meter’s cal factor array.
Example: A 500 MHz to 1 GHz Flatness Correction Array with 10 Correction Values
Create the User Flatness Array
1.
Configure the signal generator:
a.
Preset the signal generator.
b.
Set the signal generator’s connection type to the power meter/sensor:
i.
Press
AMPTD
>
More
>
User Flatness
>
Configure Power Meter
>
Connection Settings >
Connection Type
>
connection type
.
ii.
If connection type is USB:
1.
Zero Sensor
2.
Go to step
.
else
If Sockets LAN or VXI–11 LAN: Press
Power Meter IP Address
>
power meter’s or LAN–
GPIB gateway IP address
>
Enter
.
iii.
If Sockets LAN: Press
Power Meter IP Port
>
IP port
>
Enter
.
else
If VXI–11: Press
PM VXI–11 Device Name
>
device name
>
Enter
.
When connecting directly to the power meter, enter the
device name
as specified in the
power meter’s documentation. Typically, this is “
inst0
” and is case sensitive for some
power meters. Refer to your power meter’s documentation, the Keysight Connectivity
Guide USB/LAN/GPIB Connectivity Guide (E2094–90009), and Keysight X-Series FAQs
“How do I connect to the LAN?”
When connecting via a LAN–GPIB gateway, enter the SICL address of the power meter.
Typically, this is “
gpib0,13
”, where “
gpib0
” is the GPIB SICL interface name of the
gateway and “
13
” is the GPIB address of the power meter. Refer to the Keysight
Connectivity Guide USB/LAN/GPIB Connectivity Guide (E2094–90009), Keysight
X-Series FAQs “How do I connect to the LAN?”, and to the E5810A User’s Guide or
equivalent, LAN/GPIB gateway device.
The signal generator’s RF Output LED remains unchanged during zeroing
of the power sensor (e.g. if the RF Output LED was on prior to starting the
Zeroing of the power sensor, the LED remains on throughout the
zero/calibration). But, actually the instrument’s firmware has turned off
the RF Output’s power.
For operating information on your particular power meter/sensor, refer to
its operating guide.