52
Keysight EXG and MXG X-Series Signal Generators User’s Guide
Basic Operation
Configuring a Swept Output
13.
As desired, repeat step
for the remaining points for which you want to select a waveform.
The following figure shows an example of how this might look.
14.
Turn sweep on:
Press
Return
>
Return
>
Return
>
Sweep
>
Freq Off On
>
Amptd Off On
>
Waveform Off On
.
15.
If it is not already on, turn the RF output on:
Press
RF On/Off
.
The
SWEEP
annunciator appears on the display, indicating that the signal generator is
sweeping, and the progress bar shows the progression of the sweep.
Example: Using a Single Sweep
1.
Set up either a step sweep (
) or a list sweep (
).
2.
In the List/Sweep menu, set the sweep repeat to single:
Press
Sweep Repeat Single Cont
to highlight Single.
Sweep does not occur until you trigger it.
Note that the
WINIT
annunciator appears on the display, indicating that the sweep is waiting
to be initiated.
3.
If not already on, turn the RF output on: Press
RF On/Off
.
4.
Initiate the sweep: Press
Single Sweep
.
A single repetition of the configured sweep is available at the RF Output connector.
As the signal generator sweeps, the
SWEEP
annunciator replaces
WINIT
on the display, and the
progress bar shows the progression of the sweep.
At the end of the sweep, there is no progress bar, and the
WINIT
annunciator replaces
SWEEP
.
If the instrument is in manual sweep (
), the active row (row 6 in the
figure above) is the selected (manual) point, and the signal generator
outputs the settings for that selection when the RF output is on.
The empty entry is equivalent to
choosing CW (no modulation).