336
Keysight EXG and MXG X-Series Signal Generators User’s Guide
BERT (Option UN7)
Verifying BERT Operation
Measurement Setup Using Self-Test Mode
The following steps set up the signal generator for the BERT measurement selt-test.
1.
Refer to
and make the following connections on the signal generator’s rear panel.
—
DATA OUT
(
Aux I/O
connector pin 15) to BER DATA IN (BNC connector labeled EVENT 1).
—
DATA CLK OUT
(
Aux I/O
connector pin 17) to BER CLK IN (BNC connector labeled BB
TRIG1).
Figure 12-16
Rear Panel Connectors for BERT Configuration
2.
Press the
Preset
hardkey. This configures the signal generator to a pre-defined state.
3.
Press the
Aux Fctn
hardkey.
4.
Press
Digital Bus Connectivity PXB/BERT N5102A to PXB/BERT > Confirm Changes > BERT >
BERT Off On
to On >
BER Display > BER Display % Exp
to % >
Display Update Cycle End Cont
to Cont.
5.
Press
BERT > I/O Setup > AUX I/O Out
>
PN9 Out
.
6.
Press
Return > Configure Trigger > BERT Trigger
>
Immediate
.
7.
Press
Return > BERT > On.
BER Data In
BER Gate In
BER Clock In
BER
Me
as
En
d
BE
R Sy
nc Loss
BE
R
T
es
t Ou
t
BE
R
Ga
te
Out
BER
No D
ata
AU
X I
/O
GN
D
Содержание N5171B EXG
Страница 1: ...User s Guide Keysight X Series Signal Generators N5171B 72B 73B EXG N5181B 82B 83B MXG...
Страница 4: ...iv...
Страница 356: ...340 Keysight EXG and MXG X Series Signal Generators User s Guide BERT Option UN7 Verifying BERT Operation...
Страница 452: ...436 Keysight EXG and MXG X Series Signal Generators User s Guide Troubleshooting Contacting Keysight Technologies...