![Keysight Technologies N5171B EXG Скачать руководство пользователя страница 69](http://html1.mh-extra.com/html/keysight-technologies/n5171b-exg/n5171b-exg_user-manual_1962194069.webp)
Keysight EXG and MXG X-Series Signal Generators User’s Guide
53
Basic Operation
Configuring a Swept Output
Example: Manual Control of Sweep
1.
Set up either a step sweep (
) or a list sweep (
).
2.
In the Sweep/List menu, select a parameter to sweep: Press
Sweep
>
parameter
>
Return
.
3.
Select manual mode: Press
More
>
Manual Mode Off On
.
When you select manual mode, the current sweep point becomes the selected manual point.
4.
If it is not already on, turn the RF output on: Press
RF On/Off
.
5.
Select the desired point to output: Press
Manual Point
>
number
>
Enter
.
The progress bar changes to indicate the selected point.
6.
Use the knob or arrow keys to move from point to point. As you select each point, the RF
output changes to the settings in that selection.
The parameters of the selected sweep point (point 3
of 6 in this example) define the signal available at the
RF Output connector.
When you enter a manual point,
the progress bar moves to and
stops at the selected point.
The
SWMAN
annunciator
indicates that the sweep
is in manual mode.
When you turn manual mode on, the
current sweep point becomes the
manual point.
The currently sweeping
parameter stops and
displays the selected point.
Содержание N5171B EXG
Страница 1: ...User s Guide Keysight X Series Signal Generators N5171B 72B 73B EXG N5181B 82B 83B MXG...
Страница 4: ...iv...
Страница 356: ...340 Keysight EXG and MXG X Series Signal Generators User s Guide BERT Option UN7 Verifying BERT Operation...
Страница 452: ...436 Keysight EXG and MXG X Series Signal Generators User s Guide Troubleshooting Contacting Keysight Technologies...