108
Keysight EXG and MXG X-Series Signal Generators User’s Guide
Optimize Performance
Using an Output Offset, Reference, or Multiplier
To set a new frequency or amplitude reference, turn the frequency reference off, and then follow the
steps above.
Setting a Frequency Multiplier
Using a frequency multiplier, the signal generator can display a frequency that is the multiple
(positive or negative) of the output value.
Displayed Value = multiplier value ´ output frequency
Output Frequency = displayed value ¸ multiplier value
To set a frequency multiplier:
1.
Press
Frequency
>
Freq Multiplier
>
multiplier value
>
x
.
2.
Set the desired frequency.
The display equals the output frequency times the multiplier value.
Examples
Parameter
Example
#1
Example
#2
Example
#3
Comments
Frequency Multiplier:
3
−
3
4
The multiplier range can be set from:
+0.001 to +1000
–1000 to –0.001
Entered (and displayed)
Value:
600 MHz
−
600
MHz
8 GHz
Output Frequency:
200 MHz
200 MHz
2 GHz
The signal generator alerts you if the output frequency is
out of range.
Indicates that a frequency multiplier is on
Содержание N5171B EXG