Keysight EXG and MXG X-Series Signal Generators User’s Guide
207
Basic Digital Operation (Option 653/655/656/657)
Using Waveform Markers
Sample range begins on first point of signal
Negative range set between signal and
off time
Example of Incorrect Use
Waveform: 1022 points
Marker range: 110–1022
Marker polarity: Negative
This figure shows that a negative polarity marker goes
low during the marker
on
points; the marker signal goes
high during the
off
points. The ALC samples the
waveform during the
off
marker points.
Marker On
Marker On
Sampling both on and off time sets the modulator
circuitry incorrectly for higher signal levels. Note the
increased amplitude at the beginning of the pulse.
Marker
Off
Marker On
Marker
Marker
Off
On