
Keysight EXG and MXG X-Series Signal Generators User’s Guide
101
Optimize Performance
Using External Leveling (N5173B/83B Only)
To remove the error message, press the down arrow key until the message is gone. The
error appears when an attempt is made to increase the amplitude beyond the maximum
value as it relates to the current attenuator setting.
Adjusting the Signal Generator Display’s Amplitude Value
When using external leveling, the signal generator’s displayed amplitude value will not match the
leveled power of the signal at the output of the coupler/splitter. To compensate for this difference,
the signal generator provides two methods for configuring the displayed power value so that it
closely matches the measured value at the output of the coupler/splitter.
1.
Connect and configure a measurement instrument:
a.
Connect the output of the coupler/splitter to either a power meter or a signal analyzer.
b.
Configure the power meter/signal analyzer for measuring the power level of the signal.
2.
Adjust the signal generator’s displayed amplitude value:
—
If using the
Ext Leveling Amptd Offset
Softkey:
This softkey uses 16 dB as its zero reference. The 16dB is the coupling factor of the internal
detector.
a.
On the signal generator, press
AMPTD
>
Leveling Control
>
Ext Leveling Amptd Offset
.
b.
While viewing the carrier amplitude value on the signal generator display, use the RPG
knob (detent knob) to adjust the offset value until the integer part of the displayed
amplitude value is the same as the integer portion of the measured value.
Each detent position adjusts the value by 1 dB.
c.
Using the number keypad, make the necessary fractional adjustments to the display
amplitude value.
—
If using the
Amptd Offset Softkey
:
a.
On the signal generator, press
AMPTD
>
More
>
Amptd Offset
.
b.
Calculate the difference between the signal generators displayed Amplitude value and
the measured value.
c.
Using the numeric keypad, enter this difference as the
Amptd Offset
softkey value.
Содержание N5171B EXG
Страница 1: ...User s Guide Keysight X Series Signal Generators N5171B 72B 73B EXG N5181B 82B 83B MXG...
Страница 4: ...iv...
Страница 356: ...340 Keysight EXG and MXG X Series Signal Generators User s Guide BERT Option UN7 Verifying BERT Operation...
Страница 452: ...436 Keysight EXG and MXG X Series Signal Generators User s Guide Troubleshooting Contacting Keysight Technologies...