Series 2600B System SourceMeter® Instrument Reference Manual
Section 7: TSP command reference
2600BS-901-01 Rev. B / May 2013
7-143
Details
When set to
1
, the IEEE Std 488.2 common commands generate prompts if prompting is enabled with the
localnode.prompts
attribute. If set to
1
, limit the number of
*trg
commands sent to a running script to 50
regardless of the setting of the
localnode.prompts
attribute.
When set to
0
, IEEE Std 488.2 common commands will not generate prompts. When using the
*trg
command
with a script that executes
trigger.wait()
repeatedly, set
localnode.prompts4882
to
0
to avoid
problems associated with the command interface input queue filling.
This attribute resets to the default value each time the instrument power is cycled.
When using this command from a remote node,
localnode
should be replaced with the node reference, for
example
node[5].prompts4882
.
Example
localnode.prompts4882 = 0
Disables IEEE Std 488.2 common command prompting.
Also see
localnode.prompts
(on page 7-141)
localnode.reset()
This function resets the local node instrument.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
localnode.reset()
Details
If you want to reset a specific instrument or a subordinate node, use the
node[
X
].reset()
command.
A local node reset includes:
•
Source-measure unit (SMU) attributes affected by a SMU reset are reset
•
Other settings are restored back to factory default settings
A
localnode.reset()
is different than a
reset()
because
reset()
resets the entire system.
When using this command from a remote node,
localnode
should be replaced with the node reference, for
example
node[5].reset()
.
Example
localnode.reset()
Resets the local node.
Also see
reset()
(on page 7-160)
smuX.reset()
(on page 7-216)
localnode.revision
This attribute stores the firmware revision level.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (R)
Yes
Not applicable
Not applicable
Not applicable