Section 7: TSP command reference
Series 2600B System SourceMeter® Instrument Reference Manual
7-158
2600BS-901-01 Rev. B / May 2013
Usage
PulseVMeasureI(
smu, bias, level, ton, toff, points
)
smu
System SourceMeter
®
instrument channel (for example,
smua
refers to SMU
channel A)
bias
Bias level in volts
level
Pulse level in volts
ton
Pulse on time in seconds
toff
Pulse off time in seconds
points
Number of pulse-measure cycles
Details
If any parameters are omitted or
nil
, the operator is prompted to enter them using the front panel. Data for
pulsed current measurements, voltage levels, and timestamps are stored in
smu
X
.nvbuffer1
.
To perform the specified number of pulse V, measure I cycles, this function:
1. Sets the
smu
to output
bias
volts and dwell for
toff
seconds
2. Sets the
smu
to output
level
volts and dwell for
ton
seconds
3. Performs voltage measurement with source at
level
volts
4. Sets the
smu
to output
bias
volts for
toff
seconds
5. Repeats steps 2 through 4 for the remaining
points
pulse-measure cycles
Figure 135: PulseVMeasureI()
Example
smua.measure.nplc = 0.001
PulseVMeasureI(smua, -1, 1, 1E-3, 2E-3, 20)
SMU A outputs
−
1 V and dwells for 2 ms,
outputs 1 V and dwells for 1 ms. The current
measurements occur during each 1 ms
dwell period. After the measurement, the
output returns to
−
1 V and dwells for 2 ms.
This pulse-measure process repeats 19
more times.
Also see
KIPulse factory script
(on page 5-21)
QueryPulseConfig()
This KIPulse factory script function allows you to inspect the settings of the preconfigured pulse train assigned to
tag
.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
No