Section 8: Troubleshooting guide
Series 2600B System SourceMeter® Instrument Reference Manual
8-6
2600BS-901-01 Rev. B / May 2013
Error summary
Error number
Error level
Error Message
5010
SERIOUS
No such SMU function
5011
RECOVERABLE
Operation not permitted while cal is locked
5012
RECOVERABLE
Cal data not saved - save or restore before lock
5013
RECOVERABLE
Cannot save cal data - unlock before save
5014
RECOVERABLE
Cannot restore cal data - unlock before restore
5015
RECOVERABLE
Save to cal set disallowed
5016
RECOVERABLE
Cannot change cal date - unlock before operation
5017
RECOVERABLE
Cannot change cal constants - unlock before operation
5018
SERIOUS
Cal version inconsistency
5019
RECOVERABLE
Cannot unlock - invalid password
5021
SERIOUS
Cannot restore default calset. Using previous calset
5022
SERIOUS
Cannot restore previous calset. Using factory calset
5023
SERIOUS
Cannot restore factory calset. Using nominal calset
5024
SERIOUS
Cannot restore nominal calset. Using firmware defaults
5025
RECOVERABLE
Cannot set filter.count > 1 when measure.count > 1
5027
RECOVERABLE
Unlock cal data with factory password
5028
RECOVERABLE
Cannot perform requested operation while source autorange is
enabled
5029
RECOVERABLE
Cannot save without changing cal adjustment date
5032
RECOVERABLE
Cannot change this setting unless buffer is cleared
5033
RECOVERABLE
Reading buffer not found within device
5038
RECOVERABLE
Index exceeds maximum reading
5040
RECOVERABLE
Cannot use same reading buffer for multiple overlapped
measurements
5041
SERIOUS
Output Enable not asserted
5042
RECOVERABLE
Cannot perform requested action while an overlapped operation is in
progress
5043
RECOVERABLE
Cannot perform requested operation while voltage measure
autorange is enabled
5044
RECOVERABLE
Cannot perform requested operation while current measure
autorange is enabled
5045
RECOVERABLE
Cannot perform requested operation while filter is enabled
5046
SERIOUS
SMU too hot
5047
RECOVERABLE
Minimum timestamp resolution is 1us
5048
RECOVERABLE
Contact check not valid with HIGH-Z OUTPUT off
5049
RECOVERABLE
Contact check not valid while an active current source
5050
RECOVERABLE
I limit too low for contact check
5051
FATAL
Model number/SMU hardware mismatch. Disconnect DUT and cycle
power
5052
RECOVERABLE
Interlock engaged; system stabilizing
5053
RECOVERABLE
Unstable output detected - Measurements may not be valid
5055
RECOVERABLE
Cannot change adjustment date - change cal constants before
operation
5059
RECOVERABLE
trigger.source.action enabled without configuration
5060
RECOVERABLE
trigger.measure.action enabled without configuration
5061
RECOVERABLE
Operation not permitted while OUTPUT is off
5063
RECOVERABLE
Cannot perform requested operation while measure autozero is on
5064
RECOVERABLE
Cannot use reading buffer that collects source values
5065
RECOVERABLE
I range too low for contact check
5066
RECOVERABLE
source.offlimiti too low for contact check
5069
SERIOUS
Autorange locked for HighC mode