Series 2600B System SourceMeter® Instrument Reference Manual
Section 7: TSP command reference
2600BS-901-01 Rev. B / May 2013
7-43
ConfigPulseVMeasureISweepLog()
This
KIPulse factory script
function configures a voltage pulse train with a current measurement at each point.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
No
Usage
f
,
msg
= ConfigPulseVMeasureISweepLog(
smu, bias, start, stop, limit, ton, toff,
points, buffer, tag, sync_in, sync_out, sync_in_timeout, sync_in_abort
)
f
,
msg
= ConfigPulseVMeasureISweepLog(
smu, bias, start, stop, limit, ton, toff,
points, buffer, tag, sync_in, sync_out, sync_in_timeout
)
f
,
msg
= ConfigPulseVMeasureISweepLog(
smu, bias, start, stop, limit, ton, toff,
points, buffer, tag, sync_in, sync_out
)
f
,
msg
= ConfigPulseVMeasureISweepLog(
smu, bias, start, stop, limit, ton, toff,
points, buffer, tag, sync_in
)
f
,
msg
= ConfigPulseVMeasureISweepLog(
smu, bias, start, stop, limit, ton, toff,
points, buffer, tag
)
f
A Boolean flag; this flag is
true
when the pulse was successfully configured,
false
when errors were encountered
msg
A string message; if the
f
flag is
false
,
msg
contains an error message; if it is
true
,
msg
contains a string indicating successful configuration
smu
System SourceMeter
®
instrument channel (for example,
smua
refers to SMU
channel A)
bias
Bias level in volts
start
Pulse sweep start level in volts
stop
Pulse sweep stop level in volts
limit
Current limit (for example, compliance) in amperes
ton
Pulse on time in seconds
toff
Pulse off time in seconds
points
Number of pulse-measure cycles
buffer
Reading buffer where pulsed measurements will be stored; if this is
nil
when the
function is called, no measurements will be made when the pulse train is initiated
tag
Numeric identifier to be assigned to the defined pulse train
sync_in
Defines a digital I/O trigger input line; if programmed, the pulse train will wait for a
trigger input before executing each pulse
sync_out
Defines a digital I/O trigger output line; if programmed, the pulse train will generate a
trigger output immediately before the start of
ton
sync_in_timeout
Specifies the length of time (in seconds) to wait for input trigger; default value is 10 s
sync_in_abort
Specifies whether or not to abort pulse if input trigger is not received; if pulse aborts
because of a missed trigger, a timer timeout message is returned;
true
or
false
1
The KIPulse factory script provides examples of how to generate pulses and to provide a simple pulsing
interface. Pulses can be generated using the functions listed below.@Please note the following information about
the KIPulse factory script:This factory script only operates on the channels present in the instrument executing the
pulse functions. These functions will not operate correctly if you attempt to access instrument channels over the
TSPLink® interface.The KIPulse factory scripts are ge ...