Series 2600B System SourceMeter® Instrument Reference Manual
Section 3: Functions and features
2600BS-901-01 Rev. B / May 2013
3-17
Buffer status
The buffer reading status attribute includes the status information as a numeric value; see the
following table for values. For example, to access status information for the second element of SMU
channel A buffer 1, use the following command:
stat_info = smua.nvbuffer1.statuses[2]
Buffer status bits
Bit
Name
Hex value Description
B0
Reserved
0x01
Reserved for future use
B1
Overtemp
0x02
Over temperature condition
B2
AutoRangeMeas
0x04
Measure range was autoranged
B3
AutoRangeSrc
0x08
Source range was autoranged
B4
4Wire
0x10
4-wire (remote) sense mode enabled
B5
Rel
0x20
Rel applied to reading
B6
Compliance
0x40
Source function in compliance
B7
Filtered
0x80
Reading was filtered
Dynamic reading buffers
Reading buffers can also be allocated dynamically. Dynamic reading buffers are created and
allocated with the
smu
X
.makebuffer(
n
)
command, where
n
is the number of readings the buffer
can store. For example, the following command allocates a reading buffer named
mybuffer
that can
store 100 readings:
mybuffer = smua.makebuffer(100)
Allocated reading buffers can be deleted as follows:
mybuffer = nil
Dynamically allocated reading buffers can be used interchangeably with the
smu
X
.nvbuffer
Y
buffers that are described in
Dedicated reading buffer designations
(on page 3-11).