Section 7: TSP command reference
Series 2600B System SourceMeter® Instrument Reference Manual
7-254
2600BS-901-01 Rev. B / May 2013
Usage
eventID
= smu
X
.trigger.SWEEPING_EVENT_ID
eventID
The sweeping event number
X
Source-measure unit (SMU) channel (for example,
smua.trigger.SWEEPING_EVENT_ID
applies to SMU channel A)
Details
Set the stimulus of any trigger object to the value of this constant to have the trigger object respond to sweeping
events from this SMU.
Also see
Triggering
(on page 3-32)
status.condition
This attribute stores the status byte condition register.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (R)
Yes
Not applicable
Not saved
Not applicable
Usage
statusByte
= status.condition
statusByte
The status byte; a zero (0) indicates no bits set; other values indicate various bit
settings
Details
This attribute is used to read the status byte, which is returned as a numeric value. The binary equivalent of the
value of this attribute indicates which register bits are set. In the binary equivalent, the least significant bit is bit
B0, and the most significant bit is bit B7. For example, if a value of
1.02
(which is 129) is read as the
value of this register, the binary equivalent is 1000 0001. This value indicates that bit B0 and bit B7 are set.
B7
B6
B5
B4
B3
B2
B1
B0
**
>
>
>
>
>
>
*
1
0
0
0
0
0
0
1
* Least significant bit
** Most significant bit
The returned value can indicate one or more status events occurred. When an enabled status event occurs, a
summary bit is set in this register to indicate the event occurrence.
Содержание System SourceMeter 2601B
Страница 841: ......