Section 7: TSP command reference
Series 2600B System SourceMeter® Instrument Reference Manual
7-268
2600BS-901-01 Rev. B / May 2013
Bit
B7
B6
B5
B4
B3
B2
B1
B0
Binary value
0/1
0/1
0/1
0/1
0/1
0/1
0/1
0/1
Decimal
128
64
32
16
8
4
2
1
Weights
(2
7
)
(2
6
)
(2
5
)
(2
4
)
(2
3
)
(2
2
)
(2
1
)
(2
0
)
Example 1
nodeEnableRegister = status.MSB + status.OSB
status.node_enable = nodeEnableRegister
Sets the MSB and OSB bits of the
system node enable register using
constants.
Example 2
-- decimal 129 = binary 10000001
nodeEnableRegister = 129
status.node_enable = nodeEnableRegister
Sets the MSB and OSB bits of the
system node enable register using a
decimal value.
Also see
status.condition
(on page 7-254)
status.system.*
(on page 7-319)
Status byte and service request (SRQ)
status.node_event
This attribute stores the status node event register.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (R)
Yes
Not applicable
Not saved
0
Usage
nodeEventRegister
= status.node_event
nodeEventRegister
The status of the node event register; a zero (0) indicates no bits set; other values
indicate various bit settings
Details
This attribute is used to read the status node event register, which is returned as a numeric value (reading this
register returns a value). The binary equivalent of the value of this attribute indicates which register bits are set.
In the binary equivalent, the least significant bit is bit B0, and the most significant bit is bit B7. For example, if a
value of
1.02
(which is 129) is read as the value of this register, the binary equivalent is 1000 0001.
This value indicates that bit B0 and bit B7 are set.
B7
B6
B5
B4
B3
B2
B1
B0
**
>
>
>
>
>
>
*
1
0
0
0
0
0
0
1
* Least significant bit
** Most significant bit
The returned value can indicate one or more status events occurred.
Содержание System SourceMeter 2601B