Section 7: TSP command reference
Series 2600B System SourceMeter® Instrument Reference Manual
7-298
2600BS-901-01 Rev. B / May 2013
Details
These attributes are used to read or write to the operation status remote summary registers. Reading a status
register returns a value. The binary equivalent of the returned value indicates which register bits are set. The
least significant bit of the binary number is bit B0, and the most significant bit is bit B15.
For information about .condition, .enable, .event, .ntr, and .ptr registers, refer to
Status register set contents
(on
page E-1) and
Enable and transition registers
(on page E-19). The individual bits of this register are defined in
the following table.
Bit
Value
Description
B0
Not used
Not applicable.
B1
status.operation.remote.COMMAND_AVAILABLE
status.operation.remote.CAV
Set bit indicates there is a command
available in the execution queue.
Bit B1 decimal value: 2
Binary value: 0000 0000 0000 0010
B2-B10
Not used
Not applicable.
B11
status.operation.remote.PROMPTS_ENABLED
status.operation.remote.PRMPT
Set bit indicates command prompts are
enabled.
Bit B11 decimal value: 2,048
Binary value: 0000 0100 0000 0000
B12-B15
Not used
Not applicable.
As an example, to set bit B1 of the operation status remote summary enable register, set
status.operation.remote.enable = status.operation.remote.CAV
.
In addition to the above constants,
operationRegister
can be set to the numeric equivalent of the bit to set.
To set more than one bit of the register, set
operationRegister
to the sum of their decimal weights. For
example, to set bits B1 and B11, set
operationRegister
to 2,050 (which is the sum of 2 + 2,048).
Example
status.operation.remote.enable =
status.operation.remote.CAV
Uses a constant to set the CAV bit of the
operation status remote summary enable
register.
Also see
Operation Status Registers
(on page E-9)
status.operation.*
(on page 7-270)
status.operation.sweeping.*
This attribute contains the operation status sweeping summary register set.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute
- -
- -
- -
- -
.condition (R)
Yes
Not applicable
Not saved
Not applicable
.enable (RW)
Yes
Status reset
Not saved
0
.event (R)
Yes
Status reset
Not saved
0
.ntr (RW)
Yes
Status reset
Not saved
0
.ptr (RW)
Yes
Status reset
Not saved
Models 2601B/2611B/2635B:
2 (All bits set)
Models
2602B/2604B/2612B/2614B/2634B/2636B:
6 (All bits set)
Содержание System SourceMeter 2601B
Страница 841: ......