Series 2600B System SourceMeter® Instrument Reference Manual
Section 7: TSP command reference
2600BS-901-01 Rev. B / May 2013
7-215
Also see
smuX.measure.autorangeY
(on page 7-198)
smuX.measure.Y()
(on page 7-213)
smuX.source.autorangeY
smuX.trigger.source.limitY
smuX.trigger.source.linearY()
(on page 7-248)
smuX.trigger.source.listY()
(on page 7-249)
smuX.trigger.source.logY()
(on page 7-250)
Sweep Operation
(on page 3-20)
smuX.nvbufferY
This attribute contains a dedicated reading buffer.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (R)
Yes
Not applicable
See
Details
Not applicable
Usage
bufferVar
= smu
X
.nvbuffer
Y
bufferVar
The dedicated reading buffer
X
Source-measure unit (SMU) channel (for example,
smua.nvbuffer1
applies to
SMU channel A)
Y
SMU nonvolatile buffer (
1
or
2
)
Details
Each SMU channel contains two dedicated reading buffers:
smu
X
.nvbuffer1
and
smu
X
.nvbuffer2
.
All routines that return measurements can also store them in either reading buffer. Overlapped measurements
are always stored in a reading buffer. Synchronous measurements return either a single-point measurement or
can be stored in a reading buffer if passed to the measurement command.
The dedicated reading buffers can be saved to internal nonvolatile memory (to retain data between power
cycles) using the
smu
X
.savebuffer()
function.
Example
smua.measure.overlappedv(smua.nvbuffer1)
Store voltage readings from SMU channel A
into SMU channel A dedicated reading
buffer 1.
Also see
Configuring and running sweeps
(on page 3-29)
Reading buffers
(on page 3-6)
savebuffer()
(on page 7-161)
smuX.makebuffer()
(on page 7-197)
smuX.measure.overlappedY()
(on page 7-209)
smuX.savebuffer()
(on page 7-216)
smuX.trigger.measure.action
(on page 7-241)
smuX.trigger.measure.set()
(on page 7-241)
smuX.trigger.measure.stimulus
(on page 7-242)
smuX.trigger.measure.Y()