Section 7: TSP command reference
Series 2600B System SourceMeter® Instrument Reference Manual
7-326
2600BS-901-01 Rev. B / May 2013
Details
In an expanded system (TSP-Link), these attributes are used to read or write to the system summary registers.
They are set using a constant or a numeric value, but are returned as a numeric value. The binary equivalent of
the value indicates which register bits are set. In the binary equivalent, the least significant bit is bit B0, and the
most significant bit is bit B15. For example, if a value of
1.02
(which is 129) is read as the value of the
condition register, the binary equivalent is 0000 0000 1000 0001. This value indicates that bit B0 and bit B7 are
set.
B15
B14
B13
B12
B11
B10
B9
B8
B7
B6
B5
B4
B3
B2
B1
B0
**
>
>
>
>
>
>
>
>
>
>
>
>
>
>
*
0
0
0
0
0
0
0
0
1
0
0
0
0
0
0
1
* Least significant bit
** Most significant bit
For information about .condition, .enable, .event, .ntr, and .ptr registers, refer to
Status register set contents
(on
page E-1) and
Enable and transition registers
(on page E-19). The individual bits of this register are defined in
the following table.
Bit
Value
Description
B0
status.system4.EXTENSION_BIT
status.system4.EXT
Bit B0 decimal value: 1
B1
status.system4.NODE43
Bit B1 decimal value: 2
B2
status.system4.NODE44
Bit B2 decimal value: 4
B3
status.system4.NODE45
Bit B3 decimal value: 8
B4
status.system4.NODE46
Bit B4 decimal value: 16
B5
status.system4.NODE47
Bit B5 decimal value: 32
B6
status.system4.NODE48
Bit B6 decimal value: 64
B7
status.system4.NODE49
Bit B7 decimal value: 128
B8
status.system4.NODE50
Bit B8 decimal value: 256
B9
status.system4.NODE51
Bit B9 decimal value: 512
B10
status.system4.NODE52
Bit B10 decimal value: 1,024
B11
status.system4.NODE53
Bit B11 decimal value: 2,048
B12
status.system4.NODE54
Bit B12 decimal value: 4,096
B13
status.system4.NODE55
Bit B13 decimal value: 8,192
B14
status.system4.NODE56
Bit B14 decimal value: 16,384
B15
Not used
Not applicable
As an example, to set bit B0 of the system summary 4 enable register, set
status.system4.enable =
status.system4.enable.EXT
.
In addition to the above constants,
enableRegister
can be set to the numeric equivalent of the bit to set. To
set more than one bit of the register, set
enableRegister
to the sum of their decimal weights. For example, to
set bits B11 and B14, set
enableRegister
to 18,432 (which is the sum of 2,048 + 16,384).