Section 2: General operation
Series 2600B System SourceMeter® Instrument Reference Manual
2-82
2600BS-901-01 Rev. B / May 2013
The following Lua
fs
commands are not supported at this time:
fs.chmod()
fs.chown()
fs.stat()
Displayed error and status messages
During operation and programming, front-panel messages may be briefly displayed. Typical
messages are either status or error notifications (refer to the
Error summary list
(on page 8-3) for a
complete list of these messages and their meanings).
Status and error messages are held in a queue. For information about retrieving messages from
queues, refer to
Queues
(on page
). For information about error messages, refer to the
Troubleshooting guide
Range
The selected measurement range affects the accuracy of the measurements and the maximum signal
that can be measured. If the range is changed, the display may contain dashes instead of a reading
(for example, --.---- mA). This indicates that no measurement was taken using the range that is
presently selected. To update the displayed reading, trigger a measurement (if in local control, press
the TRIG key).
Available ranges
The following table lists the available source and measurement ranges for the Keithley Instruments
Series 2600B System SourceMeter
®
instrument.
Models 2601B/2602B/2604B
Models 2611B/2612B/2614B
Models 2634B/2635B/2636B
Voltage ranges
Current ranges
Voltage ranges
Current ranges
Voltage ranges
Current ranges
100 mV
100 nA
200 mV
100 nA
200 mV
100 pA
2,3
1 V
1 µA
2 V
1 µA
2 V
1 nA
6 V
10 µA
20 V
10 µA
20 V
10 nA
40 V
100 µA
200 V
100 µA
200 V
100 nA
1 mA
1 mA
1 µA
10 mA
10 mA
10 µA
100 mA
100 mA
100 µA
1 A
1 A
1 mA
3 A
1.5 A
10 mA
10 A
1
100 mA
1 A
1.5 A
1. 10 A range available only in pulse mode.
2. 100 pA range only for measurements.
3. 100 pA measurement range is not available on the Model 2634B.