Section 3: Functions and features
Series 2600B System SourceMeter® Instrument Reference Manual
3-30
2600BS-901-01 Rev. B / May 2013
Source and measurement delays
Whenever the source-measure unit (SMU) outputs a source value in a sweep, it also applies the
programmed source delay. The default source delay is zero (0) seconds. Set an additional source
delay using the
smu
X
.source.delay
attribute.
Whenever the SMU performs a measurement in a sweep, it also applies any configured
measurement delays. Use the
smu
X
.measure.delay
attribute to program a specific measurement
delay. The default measurement delay varies by measure range.
Initiating and running sweeps
To run a sweep, you must configure the number of sweep points to output and the number of sweeps
to perform. Use the trigger count to set the number of sweep points to output. Use the arm count to
set the number of times to perform the sweep. See
Triggering
(on page 3-32) for more information.
Examples:
To start a sweep, use the
smu
X
.trigger.initiate()
function. Sweeps are overlapped
operations, so you can use the
waitcomplete()
function as a way to suspend further operation
until the sweep is complete.
To sweep 15 source points:
smua.trigger.count = 15
To perform eight sweeps:
smua.trigger.arm.count = 8
Aborting a sweep
The
smu
X
.abort()
function can be used to terminate all overlapped operations on a
source-measure unit (SMU), including sweeps. It returns the SMU to the idle state of the remote
trigger model. See
Triggering
(on page 3-32) for more information.
Sweeping using factory scripts
Factory script functions that perform linear staircase, logarithmic staircase, and list sweeps are listed
in
Introduction to TSP operation
(on page 5-1). You can use the factory script functions to perform
and execute simple sweeps, or use them as examples for programming your own custom sweeps.
Front panel
To run a sweep from the front panel:
1. Press the
LOAD
key, and then select
FACTORY
.
2. Select the name of the test to run.
3. Press the
RUN
key, and then follow the display prompts to complete the test.
See
Introduction to TSP operation
(on page 5-1) for more information about using factory scripts.
Press the
RECALL
key to access sweep data stored in dedicated reading buffer 1. See
Reading
buffers
(on page 3-6) for more details about the buffer.