Section 7: TSP command reference
Series 2600B System SourceMeter® Instrument Reference Manual
7-98
2600BS-901-01 Rev. B / May 2013
gettimezone()
This function retrieves the local time zone.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
timeZone
= gettimezone()
timeZone
The local timezone of the instrument
Details
See
settimezone()
for additional details about the time zone format and a description of the fields.
timeZone
can be in either of the following formats:
•
If one parameter was used with
settimezone()
, the format used is:
GMThh:mm:ss
•
If four parameters were used with
settimezone()
, the format used is:
GMThh:mm:ssGMThh:mm:ss,Mmm.w.dw/hh:mm:ss,Mmm.w.dw/hh:mm:ss
Example
timezone = gettimezone()
Reads the value of the local timezone.
Also see
settimezone()
(on page 7-178)
gm_isweep()
This KIParlib factory script function performs a linear current sweep and calculates the transconductance (Gm) at
each point.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
No
Usage
gm_array, vbuf, ibuf
= gm_isweep(
smu, start_i, stop_i, points
)
gm_array
A Lua table containing the calculated Gm values at each point
vbuf
A reading buffer containing the measured voltage at each point
ibuf
A reading buffer containing the measured current at each point
smu
System SourceMeter
®
instrument channel (for example,
smua
refers to SMU
channel A)
start_i
Starting current level of the sweep
stop_i
Ending current level of the sweep
points
Number of measurements between
start_i
and
stop_i
(must be
≥
2)
Содержание System SourceMeter 2601B