Series 2600B System SourceMeter® Instrument Reference Manual
Section 7: TSP command reference
2600BS-901-01 Rev. B / May 2013
7-253
smuX.trigger.SOURCE_COMPLETE_EVENT_ID
This constant contains the source complete event number.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Constant
Yes
Usage
eventID
= smu
X
.trigger.SOURCE_COMPLETE_EVENT_ID
eventID
The source action complete event number
X
Source-measure unit (SMU) channel (for example,
smua.trigger.SOURCE_COMPLETE_EVENT_ID
applies to SMU channel A)
Details
Set the stimulus of any trigger object to the value of this constant to have the trigger object respond to source
complete events from this source-measure unit (SMU).
Also see
Triggering
(on page 3-32)
smuX.trigger.SWEEP_COMPLETE_EVENT_ID
This constant contains the sweep complete event number.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Constant
Yes
Usage
eventID
= smu
X
.trigger.SWEEP_COMPLETE_EVENT_ID
eventID
The sweep complete event number
X
Source-measure unit (SMU) channel (for example,
smua.trigger.SWEEP_COMPLETE_EVENT_ID
applies to SMU channel A)
Details
Set the stimulus of any trigger object to the value of this constant to have the trigger object respond to sweep
complete events from this SMU.
Also see
Triggering
(on page 3-32)
smuX.trigger.SWEEPING_EVENT_ID
This constant contains the sweeping event number.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Constant
Yes
Содержание System SourceMeter 2601B