Series 2600B System SourceMeter® Instrument Reference Manual
Section 7: TSP command reference
2600BS-901-01 Rev. B / May 2013
7-29
Example 2
printbuffer(1, 6,
smua.nvbuffer1.sourceranges)
Print the source ranges that were used for the
first 6 readings stored in source-measure unit
(SMU) A, buffer 1.
Example output:
1.00000e-04, 1.00000e-04,
1.00000e-04, 1.00000e-04,
1.00000e-04, 1.00000e-04
Also see
bufferVar.measurefunctions
(on page 7-22)
bufferVar.measureranges
bufferVar.n
(on page 7-24)
bufferVar.readings
(on page 7-25)
bufferVar.sourcefunctions
bufferVar.sourceoutputstates
(on page 7-27)
bufferVar.sourcevalues
(on page 7-29)
bufferVar.statuses
(on page 7-30)
bufferVar.timestamps
(on page 7-32)
Reading buffers
(on page 3-6)
bufferVar.sourcevalues
When enabled by the
bufferVar
.collectsourcevalues
attribute, this attribute contains the source levels
being output when readings in the reading buffer were acquired.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (R)
Yes
Clearing the buffer
See
Details
Not applicable
Usage
sourcevalue
=
bufferVar
.sourcevalues[
N
]
sourcevalue
The source's output value when reading
N
of the specified buffer was acquired
bufferVar
The reading buffer; can be a dynamically allocated buffer (user-defined), or a
dedicated reading buffer (such as
smua.nvbuffer1
)
N
The reading number (
1
to
bufferVar
.n
)
Details
If the
bufferVar
.collectsourcevalues
attribute is enabled before readings are taken, the
bufferVar
.sourcevalues
buffer recall attribute is like an array (a Lua table) of the sourced value in effect at
the time of the reading. Note that you can set the
bufferVar
.collectsourcevalues
attribute only if the
affected reading buffer is empty. See
bufferVar.collectsourcevalues
(on page 7-19) for more detailed information.
For dedicated reading buffers, all buffer attributes are saved to nonvolatile memory only when the reading buffer
is saved to nonvolatile memory.
Example 1
sourcevalue = smua.nvbuffer1.sourcevalues[1]
Get the sourced value of the first reading
stored in source-measure unit (SMU) A,
dedicated reading buffer 1.
Содержание System SourceMeter 2601B