Series 2600B System SourceMeter® Instrument Reference Manual
Section 3: Functions and features
2600BS-901-01 Rev. B / May 2013
3-31
Sweep programming examples
Procedures for programming and running a sweep for three sweep types are given on the following
pages. Each of these procedures includes commands for a typical sweep example. The following
table summarizes parameters for each of these examples.
You can retrieve the source code for the factory scripts by using the
scriptVar.list()
(on page
or
scriptVar.source
(on page
) commands.
Sweep example parameters
Sweep type
Parameters for sweep examples
Linear staircase sweep
Start current: 1 mA
Stop current: 10 mA
Settling time: 0.1 s
Number of points: 10
Pulse current sweep
Bias current: 1 mA
On current: 10 mA
Pulse on time: 10 ms
Pulse off time: 50 ms
Number of points: 10
List sweep
Points: 3 V, 1 V, 4 V, 5 V, 2 V
Settling time 0.1 s
Number of points: 5
Linear staircase sweep example
The programming example below illustrates a staircase sweep.
-- Restore Series 2600B defaults.
smua.reset()
-- Set compliance to 1 V.
smua.source.limitv = 1
1. Configure source functions.
Restores defaults and sets the
compliance to 1 V.
-- Linear staircase sweep
-- 1 mA to 10 mA, 0.1 second delay,
-- 10 points.
SweepILinMeasureV(smua, 1e-3, 10e-3, 0.1, 10)
2. Configure and execute the
sweep.
Configures a linear staircase
current sweep from 1 mA to 10 mA
with 10 points and a 0.1 second
settling time.
printbuffer(1, 10, smua.nvbuffer1.readings)
3. Request readings.
Requests readings from buffer 1.