Series 2600B System SourceMeter® Instrument Reference Manual
Section 7: TSP command reference
2600BS-901-01 Rev. B / May 2013
7-373
Details
This function sets and gets the termination character sequence that is used to indicate the end of a line for a
TSP-Net connection.
Using the
termSequence
parameter sets the termination sequence. The present termination sequence is
always returned.
For the
termSequence
parameter, use the same values listed in the table above for type. There are four
possible combinations, all of which are made up of line feeds (LF or 0x10) and carriage returns (CR or 0x13). For
TSP-enabled devices, the default is
tspnet.TERM_LF
. For devices that are not TSP-enabled, the default is
tspnet.TERM_CRLF
.
Example
deviceID = tspnet.connect("192.0.2.1")
if deviceID then
tspnet.termination(deviceID,
tspnet.TERM_LF)
end
Sets termination type for IP address
192.0.2.1 to
TERM_LF
.
Also see
tspnet.connect()
(on page 7-367)
tspnet.disconnect()
(on page 7-368)
tspnet.timeout
This attribute sets the timeout value for the
tspnet.connect()
,
tspnet.execute()
, and
tspnet.read()
commands.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
No
Instrument reset
Recall setup
Not saved
20.0 (20 s)
Usage
value
= tspnet.timeout
tspnet.timeout =
value
value
The timeout duration in seconds (0.001 to 30.000)
Details
This attribute sets the amount of time the
tspnet.connect()
,
tspnet.execute()
, and
tspnet.read()
commands will wait for a response.
The time is specified in seconds. The timeout may be specified to millisecond resolution, but is only accurate to
the nearest 10 ms.
Example
tspnet.timeout = 2.0
Sets the timeout duration to two seconds.
Also see
tspnet.connect()
(on page 7-367)
tspnet.execute()
(on page 7-369)
tspnet.read()
(on page 7-370)
Содержание System SourceMeter 2601B
Страница 841: ......