![background image](http://html2.mh-extra.com/html/keithley/sourcemeter-2400/sourcemeter-2400_user-manual_4037086168.webp)
2400 Series SourceMeter
®
User’s Manual
Source-Measure Concepts
6-25
Figure 6-11
I-Source operating examples
Voltage Limit
Load Line
Operating
Point
Current Source
Load Line
I-Source (I
S
)
V-Meter
(VM)
20V
40V
100mA
200
Ω
DUT Load Line (R)
V
M
= I
S
· R
= (100mA) (200
Ω
)
= 20V
A. Normal I-Source Operation
Voltage Limit
Load Line
Operating
Point
Current Source
Load Line
I-Source (I
S
)
V-Meter
(V
M
)
40V
100mA
I
S
= V
M
/ R
= 40V / 800
Ω
= 50mA
B. I-Source In Compliance
50mA
800
Ω
DUT Load Line (R)
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Содержание SourceMeter 2400
Страница 508: ...A Specifications Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 520: ...B StatusandErrorMessages Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 530: ...C DataFlow Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 534: ...D IEEE 488BusOverview Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 554: ...F ContactCheckFunction Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 574: ...G GPIB488 1Protocol Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 589: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 591: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 592: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...