2400 Series SourceMeter
®
User’s Manual
Remote Operations
14-3
Remote-to-local transition
When changing from remote to local operation, the following actions occur.
•
The SourceMeter stops performing source-measure operations and returns to the
idle state (ARM annunciator off).
•
All sweep operations are aborted.
•
All user-defined display messages are cancelled.
•
The display is turned on (if it was previously turned off).
•
Source autoranging is disabled.
•
Concurrent measurements are enabled.
•
If resistance was enabled, source readback is enabled.
•
The display is set to the default toggle state.
•
Readings are continuously taken (if OUTPUT is on).
Selecting an interface
The SourceMeter supports two built-in remote interfaces:
•
GPIB (General Purpose Interface Bus)
•
RS-232 interface
You can use only one interface at a time. The factory interface selection is the GPIB bus.
You can select the interface only from the front panel. The interface selection is stored in
non-volatile memory; it does not change when power has been off or after a remote inter-
face reset.
The GPIB bus is the IEEE-488 interface. You must select a unique address for the Source-
Meter. The address is displayed when the instrument is turned on. At the factory, the
address is set to 24.
The RS-232 interface is a serial interface. Programmable aspects of this interface include
the following (factory default settings are shown in parentheses):
•
Baud rate (9600)
•
Data bits (8)
•
Parity (none)
•
Terminator (CR)
•
Flow control (none)
An interface is selected and configured from the COMMUNICATIONS option of the
Main Menu (
Section 1
, “
Main menu
”). For details on the programmable aspects of the
interfaces, see
“Primary address,” page 14-6
, and
“RS-232 interface operation,”
page 14-18
.
NOTE
When changing interface selections, the SourceMeter performs a power-on
reset. To check and/or change options of the selected interface, you must re-
enter the menu structure.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Содержание SourceMeter 2400
Страница 508: ...A Specifications Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 520: ...B StatusandErrorMessages Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 530: ...C DataFlow Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 534: ...D IEEE 488BusOverview Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 554: ...F ContactCheckFunction Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 574: ...G GPIB488 1Protocol Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 589: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 591: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 592: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...