SCPI Command Reference 18-121
WARNING
With auto output-off disabled, the source output will remain on after
all programmed source-measure operations are completed. Beware of
hazardous voltage (
≥
30VDC, 42.4 peak-to-peak) that may be present
on the output terminals.
With auto output-off enabled, an initiation will start operation immedi-
ately. The source output will automatically turn on at the beginning of
each SDM (source-delay-measure) cycle and turn off after each mea-
surement is completed.
Abort source/measure cycle
ABORt
Abort operation
Description
When this action command is sent, the SourceMeter aborts operation
and returns to the idle state.
A faster way to return to idle is to use the DCL or SDC command.
With auto output-off enabled (:SOURce1:CLEar:AUTO ON), the output
will remain on if operation is terminated before the output has a chance
to automatically turn off.
Program trigger model
COUNt <n>
:ARM[:SEQuence[1]][LAYer[1]]:COUNt <n>
Set arm count
:TRIGger[:SEQuence[1]]:COUNt <n>
Set trigger count
Parameters
<n> =
1 to 2500
Specify count (see NOTE)
DEFault
Sets count to 1
MINimum
Sets count to 1
MAXimum
See NOTE
INFinite
(ARM:COUNt only)
NOTE
The product of arm count and trigger count cannot exceed 2500.
Query
:COUNt?
Queries programmed count
:COUNt? DEFault
Queries *RST default count
:COUNt? MINimum
Queries lowest allowable count
:COUNt? MAXimum
Queries largest allowable count
Description
This command is used to specify how many times an operation is per-
formed in the specified layer of the trigger model.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Содержание SourceMeter 2400
Страница 508: ...A Specifications Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 520: ...B StatusandErrorMessages Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 530: ...C DataFlow Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 534: ...D IEEE 488BusOverview Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 554: ...F ContactCheckFunction Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 574: ...G GPIB488 1Protocol Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 589: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 591: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 592: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...