2400 Series SourceMeter
®
User’s Manual
SCPI Command Reference 18-125
Description
When TLINk is the selected Trigger Layer control source, and an event
detector in the Trigger Layer is enabled, operation will hold up at that
detector until an input trigger is received via the Trigger Link. When the
event detector is disabled, operation will not hold up. It continues and
performs the appropriate action.
A Trigger Layer event detector is enabled by including the parameter
name in the event list for the INPut command. For example, to enable
the Source Event Detector and Measure Event Detector, send the fol-
lowing command:
:TRIGger:INPut SOURce, SENSe
The Delay Event Detector will be disabled since the DELay parameter
is not included in the above event list.
NOTE
To disable all the Trigger Layer event detectors, the NONE parameter must be
sent alone (i.e. trigger:input none). If it is listed with any of the other parame-
ters, NONE will be ignored.
ILINe <NRf>
:ARM[:SEQuence[1]][LAYer[1]][:TCONfigure]:ILINe <NRf>
Select input line; arm layer
:TRIGger[:SEQuence[1]][:TCONfigure]:ILINe <NRf>
Select input line; trigger layer
Parameters
<NRf> =
1
Line #1
2
Line #2
3
Line #3
4
Line #4
Query
:ILINe?
Query input trigger line
Description
This command is used to select input lines for the Trigger Link. For nor-
mal operation, Trigger Link input and output (see “
OLINe <NRf>
”)
should not share the same line.
OLINe <NRf>
:ARM[:SEQuence[1]][LAYer[1]][:TCONfigure]:OLINe <NRf>
Select output line; arm layer
:TRIGger[:SEQuence[1]][:TCONfigure]:OLINe <NRf>
Select output line; trigger layer
Parameters
<NRf> =
1
Line #1
2
Line #2
3
Line #3
4
Line #4
Query
:OLINe?
Query output trigger line
Description
This command is used to select output lines for the Trigger Link. For
normal operation, Trigger Link input and output (see “
ILINe <NRf>
”)
should not share the same line.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Содержание SourceMeter 2400
Страница 508: ...A Specifications Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 520: ...B StatusandErrorMessages Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 530: ...C DataFlow Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 534: ...D IEEE 488BusOverview Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 554: ...F ContactCheckFunction Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 574: ...G GPIB488 1Protocol Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 589: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 591: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 592: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...