![background image](http://html2.mh-extra.com/html/keithley/sourcemeter-2400/sourcemeter-2400_user-manual_4037086470.webp)
2400 Series SourceMeter
®
User’s Manual
SCPI Command Reference
18-89
2420
<n> =
-6.3 to 6.3
Set I-Source level (amps)
-128 to 128
Set V-Source level (volts)
DEFault
0A or 0V
MINimum
-6.3A or -128V
MAXimum
+6.3A or +128V
2425 and 2430 DC Mode
<n> =
-6.3 to 6.3
Set I-Source level (amps)
-210 to 210
Set V-Source level (volts)
DEFault
0A or 0V
MINimum
-6.3A or -210V
MAXimum
+6.3A or +210V
2430 Pulse Mode
<n> =
-21 to 21
Set I-Source level (amps)
-210 to 210
Set V-Source level (volts)
DEFault
0A or 0V
MINimum
-21A or -210V
MAXimum
+21A or +210V
2440
<n> =
-10.5 to 10.5
Set I-Source level (amps)
-42 to 42
Set V-Source level (volts)
DEFault
0A or 0V
MINimum
-10.5A or -84V
MAXimum
+10.5A or +84V
Query
:STEP
Query step size for sweep
:STEP? DEFault
Query *RST default level
:STEP? MINimum
Query lowest allowable level
:STEP? MAXimum
Query highest allowable level
Description
This command is used to specify a step size for a linear sweep. When
the sweep is started, the source level changes in equal steps from the
start level to the stop level. A measurement is performed at each source
step (including the start and stop levels).
NOTE
This command cannot be used for a logarithmic sweep. Use the :POINTs com-
mand to set the source-measure points for a log sweep.
To avoid a setting conflicts error, make sure the step size is greater than
the start value and less than the stop value.
The number of source-measure points in a linear sweep can be calcu-
lated as follows:
Points = [(Stop - Start) / Step] + 1|
Points = (Span / Step) + 1
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Содержание SourceMeter 2400
Страница 508: ...A Specifications Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 520: ...B StatusandErrorMessages Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 530: ...C DataFlow Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 534: ...D IEEE 488BusOverview Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 554: ...F ContactCheckFunction Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 574: ...G GPIB488 1Protocol Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 589: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 591: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 592: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...