A-10
Specifications
2400 Series SourceMeter
®
User’s Manual
Case IV: Auto-Zero disabled and all measurements disabled
Figure
A
-4
Case IV timing diagram
Trigger Latency:
225µsec max
Source Configuration: 50µsec
max
Firmware Overhead:
310µsec for Source V
590µsec for Source I
Source On Time
≅
Source Configu Source Delay + Firmware Overhead
Example:
Source Delay = 0
Source On Time
≅
50µsec + 0 + 125µsec
≅
360µsec for Source V
≅
640µsec for Source I
Trigger
Event
Source
On
Source On Time
Source
Off
Trigger
Delay
Trigger
Latency
Source
Configuration
Source
Delay
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Содержание SourceMeter 2400
Страница 508: ...A Specifications Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 520: ...B StatusandErrorMessages Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 530: ...C DataFlow Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 534: ...D IEEE 488BusOverview Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 554: ...F ContactCheckFunction Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 574: ...G GPIB488 1Protocol Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 589: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 591: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 592: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...