6-18
Source-Measure Concepts
2400 Series SourceMeter
®
User’s Manual
Models 2425 and 2430 SourceMeters
The general operating boundaries for the Models 2425 and 2430 are shown in
Figure 6-7
.
The boundaries for the Model 2425 and Model 2430 DC Mode are shown in
Figure 6-7A
,
and the boundaries for the Model 2430 Pulse Mode are shown in
Figure 6-7B
.
Model 2425 and Model 2430 DC Mode
In
Figure 6-7A
for the Model 2425 and Model 2430 DC Mode, the 3A, 20V and 1A, 100V
magnitudes are nominal values. The actual maximum output magnitudes of the
SourceMeter are 3.15A, 21V and 1.05A, 105V.
These operating boundaries are valid only if the SourceMeter is being operated in an envi-
ronment where the ambient temperature is 30˚C or less.
NOTE
Above 30˚C, high power operation could overheat the SourceMeter, causing the
output to turn off. See
“Overheating protection,” page 6-6
, and
“Power equa-
tions to avoid shutdown,” page 6-7
, for details.
The heavy solid lines show the limits for continuous output operation. Note that in quad-
rants II and IV (sink operation), the limits for the 3A and 1A ranges are derated as follows:
3A Range – Limits linearly derated from:
-3A, 0V to -2A, 20V
3A, 0V to 2A, -20V
1A Range – Limits linearly derated from:
-1A, 0V to -0.5A, 100V
1A, 0V to 0.5A, -100V
If the output duty cycle is decreased to 60% or less, sink operation limits are restored to
normal as shown by the dotted lines in
Figure 6-7
.
Model 2430 Pulse Mode
In
Figure 6-7B
for the Model 2430 Pulse Mode, the 10A and 100V magnitudes are nomi-
nal values. The actual maximum output magnitudes of the SourceMeter are 10.5A and
105V.
These operating boundaries are valid only if the SourceMeter is being operated in an envi-
ronment where the ambient temperature is 30˚C or less.
NOTE
Above 30˚C, high power operation could overheat the SourceMeter, causing the
output to turn off.
The heavy solid lines show the limits for an output duty cycle of 8%. Note that in quad-
rants II and IV (sink operation), the limits for the 10A range are derated as follows:
10A Range – Limits linearly derated from:
-10A, 0V to -6A, 100V
10A, 0V to 6A, -100V
If the output duty cycle is decreased to 5% or less, sink operation limits are restored to
normal as shown by the dotted lines in
Figure 6-7B
.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Содержание SourceMeter 2400
Страница 508: ...A Specifications Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 520: ...B StatusandErrorMessages Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 530: ...C DataFlow Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 534: ...D IEEE 488BusOverview Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 554: ...F ContactCheckFunction Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 574: ...G GPIB488 1Protocol Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 589: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 591: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 592: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...