18-46
SCPI Command Reference
2400 Series SourceMeter
®
User’s Manual
DISPlay subsystem
The display subsystem controls the display of the SourceMeter and is summarized in
Table 18-2
.
Control display
DIGits <n>
:DISPlay:DIGits <n>
Set display resolution
Parameters
<n> =
4
3.5 digit resolution
5
4.5 digit resolution
6
5.5 digit resolution
7
6.5 digit resolution
DEFault
5.5 digit resolution
MINimum
3.5 digit resolution
MAXimum
6.5 digit resolution
Query
:DIGits?
Query display resolution
:DIGits? DEFault
Query *RST default resolution
:DIGits? MINimum
Query lowest allowable display resolution
:DIGits? MAXimum
Query largest allowable display resolution
Description
This command is used to set the display resolution. Note that you can
instead use rational numbers. For example, to select 4.5 digit resolution,
you can send a parameter value of 4.5 (instead of 5). The SourceMeter
rounds the rational number to an integer.
ENABle <b>
:DISPlay:ENABle <b>
Control display circuitry
Parameters
<b> =
0 or OFF
Disable display circuitry
1 or ON
Enable display circuitry
Query
:ENABle?
Query state of display
Description
This command is used to enable and disable the front panel display cir-
cuitry. When disabled, the instrument operates at a higher speed. While
disabled, the display is frozen with the following message:
FRONT PANEL DISABLED
Press LOCAL to resume.
As reported by the message, all front panel controls (except LOCAL
and OUTPUT OFF) are disabled. Normal display operation can be
resumed by using the :ENABle command to enable the display or by
putting the SourceMeter into local.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Содержание SourceMeter 2400
Страница 508: ...A Specifications Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 520: ...B StatusandErrorMessages Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 530: ...C DataFlow Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 534: ...D IEEE 488BusOverview Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 554: ...F ContactCheckFunction Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 574: ...G GPIB488 1Protocol Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 589: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 591: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 592: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...