18-58
SCPI Command Reference
2400 Series SourceMeter
®
User’s Manual
Interlock control
STATe <b>
:OUTPut[1]:INTerlock:STATe <b>
Control hardware interlock
Parameters
<b> =
0 or OFF
Disable interlock
1 or ON
Enable interlock
Query
:STATe?
Query state of interlock
Description
This command is used to enable or disable the hardware interlock.
When enabled, the source cannot be turned on unless the interlock line
(pin 8 of the rear panel Interlock - Digital I/O connector) is pulled to a
logic low state. When the interlock line goes to a logic high state, the
source turns off. See
Section 13
, “
Digital I/O port
” and “
Safety inter-
lock
,” for details about using interlock with a test fixture.
When disabled, the logic level on the interlock line has no effect on the
output state of the source.
TRIPped?
:OUTPut[1]:INTerlock:TRIPped?
Description
This query command is used to determine if the enabled interlock has
been tripped. The tripped condition (“1”) means that the source can be
turned on (interlock line at logic low level).
A “0” will be returned if the source cannot be turned on (interlock line
at logic high level).
Output-off states
SMODe
:OUTPut[1]:SMODe <name>
Select output-off mode
Parameters
<name> =
HIMPedance
Disconnect Input/Output
NORMal
Normal output-off state
ZERO
Zero output-off state
GUARd
Guard output-off state
Query
:SMODe?
Query output off mode
Description
This command is used to select the output-off state of the SourceMeter.
With HIMPedance selected, the output relay opens when the source is
turned off. This disconnects external circuitry from the SourceMeter
Input/Output.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Содержание SourceMeter 2400
Страница 508: ...A Specifications Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 520: ...B StatusandErrorMessages Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 530: ...C DataFlow Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 534: ...D IEEE 488BusOverview Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 554: ...F ContactCheckFunction Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 574: ...G GPIB488 1Protocol Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 589: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 591: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 592: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...