![background image](http://html2.mh-extra.com/html/keithley/sourcemeter-2400/sourcemeter-2400_user-manual_4037086455.webp)
18-74
SCPI Command Reference
2400 Series SourceMeter
®
User’s Manual
[:MODE] <name>
:SOURce[1]:FUNCtion[:MODE] <name>
Select source mode
Parameters
<name> =
VOLTage
Select voltage mode
CURRent
Select current mode
MEMory
Select memory mode
Query
[:MODE]?
Query selected source
Description
This command is used to select the source mode. With VOLTage
selected, the V-Source will be used, and with CURRent selected, the
I-Source will be used.
With MEMory selected, a memory sweep can be performed. Operating
setups (up to 100) saved in memory can be sequentially recalled. This
allows multiple source/measure functions to be used in a sweep.
Select sourcing mode
MODE <name>
:SOURce[1]:CURRent:MODE <name>
Select DC sourcing mode for I-Source
:SOURce[1]:VOLTage:MODE <name>
Select DC sourcing mode for V-Source
Parameters
<name> =
FIXed
Select fixed sourcing mode
LIST
Select list sourcing mode
SWEep
Select sweep sourcing mode
Query
:MODE?
Query DC sourcing mode
Description
This command is used to select the DC sourcing mode for the specified
source. The three modes are explained as follows:
FIXed
— In this DC sourcing mode, the specified source will output a
fixed level. Use the :RANGe and :AMPLitude commands to specify the
fixed source level. (See
“Select range,” page 18-75
, and
“Set amplitude
for fixed source,” page 18-77
.)
LIST
— In this mode, the source will output levels that are specified in
a list. See “
Configure list
” for commands to define and control the exe-
cution of the list.
SWEep
— In this mode, the source will perform a voltage, current or
memory sweep. See
“Configure voltage and current sweeps,”
page 18-84
, and
“Configure memory sweep,” page 18-93
, for com-
mands to define the sweep.
NOTE
The sourcing mode will default to FIXed whenever the SourceMeter goes to the
local state.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Содержание SourceMeter 2400
Страница 508: ...A Specifications Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 520: ...B StatusandErrorMessages Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 530: ...C DataFlow Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 534: ...D IEEE 488BusOverview Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 554: ...F ContactCheckFunction Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 574: ...G GPIB488 1Protocol Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 589: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 591: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 592: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...