The MINimum and MAXimum parameters are only valid if the highest
source range is presently selected. Sending the MINimum or MAXi-
mum parameters on a lower source range will generate error -221 (Set-
ting Conflict).
TRIGgered[:AMPLitude] <n>
:SOURce[1]:CURRent[:LEVel]:TRIGgered[:AMPLitude] <n>
Set fixed I-Source amplitude when
triggered
:SOURce[1]:VOLTage[:LEVel]:TRIGgered[:AMPLitude] <n>
Set fixed V-Source amplitude when
triggered
Parameters
2400
<n> =
-1.05 to 1.05
Set I-Source amplitude (amps)
-210 to 210
Set V-Source amplitude (volts)
DEFault
0A or 0V
MINimum
-1.05A or -210V
MAXimum
+1.05A or +210V
2410
<n> =
-1.05 to 1.05
Set I-Source amplitude (amps)
-1100 to 1100
Set V-Source amplitude (volts)
DEFault
0A or 0V
MINimum
-1.05A or -1100V
MAXimum
+1.05A or +1100V
2420
<n> =
-3.15 to 3.15
Set I-Source amplitude (amps)
-63 to 63
Set V-Source amplitude (volts)
DEFault
0A or 0V
MINimum
-3.15A or -63V
MAXimum
+3.15A or +63V
2425 and 2430 DC Mode
<n> =
-3.15 to 3.15
Set I-Source amplitude (amps)
-105 to 105
Set V-Source amplitude (volts)
DEFault
0A or 0V
MINimum
-3.15A or -105V
MAXimum
+3.15A or +105V
2430 Pulse Mode
<n> =
-10.5 to 10.5
Set I-Source amplitude (amps)
-105 to 105
Set V-Source amplitude (volts)
DEFault
0A or 0V
MINimum
-10.5A or -105V
MAXimum
+10.5A or +105V
2440
<n> =
-5.25 to 5.25
Set I-Source amplitude (amps)
-42 to 42
Set V-Source amplitude (volts)
DEFault
0A or 0V
MINimum
-5.25A or -42V
MAXimum
+5.25A or +42V
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Содержание SourceMeter 2400
Страница 508: ...A Specifications Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 520: ...B StatusandErrorMessages Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 530: ...C DataFlow Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 534: ...D IEEE 488BusOverview Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 554: ...F ContactCheckFunction Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 574: ...G GPIB488 1Protocol Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 589: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 591: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 592: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...