Source operation
Figure 13-3
shows the basic output configuration for source operation. In this case, the
external relay coil is connected between the digital output line (pins 1 to 4) and ground
(pin 9). With this configuration, the digital output line must be set HI to energize the relay,
and the maximum source current is 2mA.
Figure 13-3
Source operation
Controlling digital output lines
Although the digital output lines are primarily intended for use with a device handler for
limit testing, they can also be used for other purposes such as controlling external relays or
indicator lights. You can control these lines either from the front panel or via remote as
covered below.
Front panel digital output control
Set digital output line logic levels from the front panel as follows:
1.
Press the MENU key.
2.
Select GENERAL, then press ENTER.
3.
Select DIGOUT, then press ENTER.
4.
Using the RANGE and cursor keys, set the digital output parameter to the desired
decimal value (
Table 13-1
). For example, to set the output lines to L, H, H, H, set
the digital output parameter value to 7.
5.
Press EXIT to return to normal display.
SourceMeter
External
Relay
To other
Circuits
+5V
Maximum source current: 2mA
Pin 7
Pins 1-4
Pin 9
Digital I/O
Port
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Содержание SourceMeter 2400
Страница 508: ...A Specifications Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 520: ...B StatusandErrorMessages Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 530: ...C DataFlow Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 534: ...D IEEE 488BusOverview Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 554: ...F ContactCheckFunction Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 574: ...G GPIB488 1Protocol Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 589: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 591: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 592: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...