2400 Series SourceMeter
®
User’s Manual
SCPI Command Reference 18-119
FEED <name>
:TRACe:FEED <name>
Specify readings source
Parameters
<name> =
SENSe[1]
Put raw readings in buffer
CALCulate[1]
Put Calc1 readings in buffer
CALCulate2
Put Calc2 readings in buffer
Query
:FEED?
Query buffer feed
Description
This command is used to select the source of readings to be placed in
the buffer. With SENSe[1] selected, raw readings are placed in the
buffer when storage is performed.
With CALCulate[1] selected, math expression results (Calc1) are placed
in the buffer. With CALCulate2 selected, Calc2 readings are placed in
the buffer.
TRACe:FEED cannot be changed while buffer storage is active. See
Section 9
, “
Data Store
,” for more information.
CONTrol <name>
:TRACe:FEED:CONTrol <name>
Start or stop buffer
Parameters
<name> =
NEXT
Fills buffer and stops
NEVer
Disables buffer storage
Query
:CONTrol?
Query buffer control
Description
This command is used to select the buffer control. When NEXT is
selected, the asterisk (*) annunciator turns on to indicate that the buffer
is enabled. The storage process starts when SourceMeter is taken out of
idle to perform source-measure operations.
After the buffer stores the specified number of reading arrays (as set by
the :POINTs command), the asterisk annuciator turns off to indicate that
storage is done.
With NEVer selected, storage into the buffer is disabled.
Select timestamp format
FORMat <name>
:TRACe:TSTamp:FORMat <name>
Select timestamp format
Parameters
<name> =
ABSolute
Reference to first buffer reading
DELTa
Time between buffer readings
Query
:FORMat?
Query timestamp format
Description
This command is used to select the timestamp format for buffer read-
ings. With ABSolute selected, each timestamp is referenced to the first
reading stored in the buffer. With DELTa selected, timestamps provide
the time between each buffer reading.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Содержание SourceMeter 2400
Страница 508: ...A Specifications Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 520: ...B StatusandErrorMessages Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 530: ...C DataFlow Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 534: ...D IEEE 488BusOverview Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 554: ...F ContactCheckFunction Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 574: ...G GPIB488 1Protocol Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 589: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 591: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 592: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...