2400 Series SourceMeter
®
User’s Manual
Status Structure
15-11
Status register sets
As shown in
Figure 15-1
, there are four status register sets in the status structure of the
SourceMeter; Standard Event Status, Operation Event Status, Measurement Event Status,
and Questionable Event Status.
NOTE
See
Appendix B
for details on which register bits are set by specific error and
status conditions.
Register bit descriptions
Standard Event Register
The used bits of the Standard Event Register (shown in
Figure 15-4
) are described as follows:
•
Bit B0, Operation Complete
— Set bit indicates that all pending selected device
operations are completed and the SourceMeter is ready to accept new commands.
This bit only sets in response to the *OPC? query command. See
Section 16
for
details on *OPC and *OPC?.
•
Bit B1
— Not used.
•
Bit B2, Query Error (QYE)
— Set bit indicates that you attempted to read data
from an empty Output Queue.
•
Bit B3, Device-Dependent Error (DDE)
— Set bit indicates that an instrument
operation did not execute properly due to some internal condition.
•
Bit B4, Execution Error (EXE)
— Set bit indicates that the SourceMeter detected
an error while trying to execute a command.
•
Bit B5, Command Error (CME)
— Set bit indicates that a command error has
occurred.
•
Command errors include:
•
IEEE-488.2 syntax error — SourceMeter received a message that does not fol-
low the defined syntax of the IEEE-488.2 standard.
•
Semantic error — SourceMeter received a command that was misspelled or
received an optional IEEE-488.2 command that is not implemented.
•
The instrument received a Group Execute Trigger (GET) inside a program
message.
•
Bit B6, User Request (URQ)
— Set bit indicates that the LOCAL key on the
SourceMeter front panel was pressed.
•
Bit B7, Power ON (PON)
— Set bit indicates that the SourceMeter has been
turned off and turned back on since the last time this register has been read.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Содержание SourceMeter 2400
Страница 508: ...A Specifications Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 520: ...B StatusandErrorMessages Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 530: ...C DataFlow Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 534: ...D IEEE 488BusOverview Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 554: ...F ContactCheckFunction Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 574: ...G GPIB488 1Protocol Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 589: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 591: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 592: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...