![background image](http://html2.mh-extra.com/html/keithley/sourcemeter-2400/sourcemeter-2400_user-manual_4037086480.webp)
2400 Series SourceMeter
®
User’s Manual
SCPI Command Reference
18-99
Soak time
SOAK <NRf>
:SOURce[1]:SOAK <NRf>
Set multiple mode soak time
Parameters
<NRf> =
soak time (s)
0.000 to 9999.999s
Query
:SOAK?
Query multiple mode soak time
Description
With SYST:RCMode set to MULTiple, SOUR:SOAK specifies the
amount of time after the first point of a sweep that the unit will sit in a
loop actively autoranging up and down to allow a multiple SourceMeter
configuration to settle. (See
“SYSTem subsystem,” page 18-106
.) This
process will be done only once per INIT, READ?, or MEAS? command.
The soak time is especially useful for low current measurements when
multiple down-range changes from the higher ranges are required.
NOTE
For the Model 2430 Pulse Mode, soak time is not used for sweeps.
Pulse Mode delays (Model 2430 only)
WIDth <n>
:SOURce[1]:PULSe:WIDTh <n>
Specify pulse width
Parameters
<n> =
0.00015 to 0.00500 Specify pulse width in seconds
MINimum
0.00015 seconds
MAXimum
0.00500 seconds
DEFault
0.00015 seconds
Query
:WIDTh?
Query pulse width
:WIDTh? DEFault
Query *RST default pulse width
:WIDTh? MINimum
Query lowest allowable pulse width
:WIDTh? MAXimum
Query highest allowable pulse width
Description
This command is used to set the pulse width for the Model 2430 Pulse
Mode. The pulse width is the time that the output remains on, including
the signal measurement.
The valid range of pulse width periods is from 0.15ms to 5ms. The pulse
width can be set to <0.15ms, however, 0.15ms is the minimum pulse
width that will occur. On the 10A range (source or measure), the pulse
width can be set up to 5ms, however, it will be limited to 2.5ms.
NOTE
See
Section 5
for details on Model 2430 Pulse Mode.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Содержание SourceMeter 2400
Страница 508: ...A Specifications Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 520: ...B StatusandErrorMessages Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 530: ...C DataFlow Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 534: ...D IEEE 488BusOverview Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 554: ...F ContactCheckFunction Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 574: ...G GPIB488 1Protocol Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 589: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 591: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 592: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...