18-92
SCPI Command Reference
2400 Series SourceMeter
®
User’s Manual
Description
These commands are used to define a list of source values (up to 100)
for the list sourcing mode of operation. When operation is started, the
instrument will sequentially source each current or voltage value in the
list. A measurement is performed at each source level.
The following command shows the proper format for defining an I-
Source list using current source values of 10mA, 130mA and 5mA:
:SOURce[1]:LIST:CURRent 0.01, 0.13, 0.005
If using manual source ranging, you can use auto range for source val-
ues that are not within a single range.
NOTE
In order to execute a source list, the selected source must be in the list sourcing
mode, and the product of the arm and trigger count should be at least the same
as the number of source points in the list.
Use the :FUNCtion:MODE command to select the current or voltage
source function. (See
“Select function modes,” page 18-73
.) Use the
:CURRent:MODE or VOLTage:MODE command to select the LIST
sourcing mode. (See
“Select sourcing mode,” page 18-74
.) The trigger
count is set using the TRIGger:COUNt command. (See
“TRIGger sub-
system,” page 18-120
.)
APPend <NRf list>
:SOURce[1]:LIST:CURRent:APPend <NRf list>
Add value(s) to I-Source list
:SOURce[1]:LIST:VOLTage:APPend <NRf list>
Add value(s) to V-Source list
Parameters
<NRf list> = NRf, NRf … NRf
2400
NRf =
-1.05 to 1.05
I-Source value
-210 to 210
V-Source value
2410
NRf =
-1.05 to 1.05
I-Source value
-1100 to 1100
V-Source value
2420
NRf =
-3.15 to 3.15
I-Source value
-63 to 63
V-Source value
2425 and 2430 DC Mode
NRf =
-3.15 to 3.15
I-Source value
-105 to 105
V-Source value
2430 Pulse Mode
NRf =
-10.5 to 10.5
I-Source value
-105 to 105
V-Source value
2440
NRf =
-5.25 to 5.25
I-Source value
-42 to 42
V-Source value
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Содержание SourceMeter 2400
Страница 508: ...A Specifications Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 520: ...B StatusandErrorMessages Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 530: ...C DataFlow Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 534: ...D IEEE 488BusOverview Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 554: ...F ContactCheckFunction Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 574: ...G GPIB488 1Protocol Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 589: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 591: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 592: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...