10-28
Sweep Operation
2400 Series SourceMeter
®
User’s Manual
Call OutputCmd(intGPIB, “:CALC2:LIM1:STAT ON”)
'Set Limit1 on
Call OutputCmd(intGPIB, “:CALC2:LIM1:COMP:FAIL IN”)'Set Fail Mode to In
Compliance
Call OutputCmd(intGPIB, “:CALC2:LIM1:COMP:SOUR2 8”)'Set Digital Output
Pattern for Compliance
Failure
Call OutputCmd(intGPIB, “:CALC2:LIM2:STAT ON”)
'Set Limit2 on
Call OutputCmd(intGPIB, “:CALC2:LIM2:UPP 2E-2”)
'Set Upper Limit to 20mA
Call OutputCmd(intGPIB, “:CALC2:LIM2:LOW 1E-3”)
'Set Lower Limit to 10mA
Call OutputCmd(intGPIB, “:CALC2:LIM2:UPP:SOUR2 2”) 'Set Digital Output
Pattern for Upper Limit
#2 Failure
Call OutputCmd(intGPIB, “:CALC2:LIM2:LOW:SOUR2 3”) 'Set Digital Output
Pattern for Lower Limit
#2 Failure
Call OutputCmd(intGPIB, “:CALC2:CLIM:PASS:SOUR2 13”)'Set Digital Output
Pattern for Limit #2
Passing
Call OutputCmd(intGPIB, “:CALC2:CLIM:MODE GRAD”)
'Set Limit Results to
Grading
Call OutputCmd(intGPIB, “:CALC2:CLIM:BCON END”)
'Set Binning Control
to End
Call OutputCmd(intGPIB, “:CALC2:CLIM:PASS:SML NEXT”) 'Set SML Pass Location
Call OutputCmd(intGPIB, “:CALC2:CLIM:FAIL:SML 100”)
'Set SML Fail Location
Call OutputCmd(intGPIB, “:SOUR:MEM:SAVE 1”)
'Save in Source Memory
Location 1.
'Setup Source Memory Location 2
'------------------------------
Call OutputCmd(intGPIB, “*RST”)
'Restore GPIB default
conditions.
Call OutputCmd(intGPIB, “:SOUR:FUNC VOLT”)
'Current Source
Function.
Call OutputCmd(intGPIB, “:SENS:FUNC ‘CURR:DC’”)
'Current Sense
Function.
Call OutputCmd(intGPIB, “:SENS:CURR:PROT .105”)
'Set 105mA Compliance
Call OutputCmd(intGPIB, “:SENS:CURR:RANGE .1”)
'Set 100mA Current
Measure Range
Call OutputCmd(intGPIB, “:SOUR:DEL 1”)
'Set Source Delay to 1
Call OutputCmd(intGPIB, “:SOUR:VOLT 10”)
'10V Source Voltage.
Call OutputCmd(intGPIB, “:CALC2:FEED CURR”)
'Send Current(A)
Readings to Buffer
Call OutputCmd(intGPIB, “:CALC2:LIM1:STAT ON”)
'Set Limit 1 on
Call OutputCmd(intGPIB, “:CALC2:LIM1:COMP:FAIL IN”)
'Set Fail Mode to In
Compliance
Call OutputCmd(intGPIB, “:CALC2:LIM1:COMP:SOUR2 5”)
'Set Digital Output
Pattern for Compli-
ance Failure
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Содержание SourceMeter 2400
Страница 508: ...A Specifications Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 520: ...B StatusandErrorMessages Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 530: ...C DataFlow Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 534: ...D IEEE 488BusOverview Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 554: ...F ContactCheckFunction Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 574: ...G GPIB488 1Protocol Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 589: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 591: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 592: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...